Device programing and SMART pass algorithm activation in subcutaneous implantable defibrillator patients: Data from a
Saverio Iacopino1, Ezio Santobuono2, Claudia Amellone3
1Arrhythmology Department, Maria Cecilia Hospital, Ravenna, Italy.
Pacing and Clinical Electrophysiology : PACE
|July 29, 2024
Summary
Subcutaneous implantable cardioverter-defibrillator (S-ICD) programming shows SMART Pass filter activation reduces inappropriate shocks and arrhythmias. The enhanced SMART Pass version further lowers deactivation rates and arrhythmia occurrences in S-ICD patients.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Subcutaneous implantable cardioverter-defibrillator (S-ICD) programming in clinical practice remains understudied.
- The activation status of the SMART Pass filter, designed to reduce inappropriate shocks, is a key area of investigation.
Purpose of the Study:
- To assess device programming during follow-up in S-ICD recipients.
- To evaluate the rate of detected arrhythmias in S-ICD patients.
Main Methods:
- Analysis of data from 670 S-ICD patients across 17 Italian centers, followed remotely for a median of 31 months.
- Investigation of the impact of the enhanced SMART Pass version, introduced in October 2022, on unintentional deactivation rates.
Main Results:
- SMART Pass was enabled in 87% of patients, with median conditional and shock zone cut-offs at 210 bpm and 250 bpm, respectively.
- SMART Pass activation was associated with significantly lower rates of shocks (0.12 vs. 0.20/patient-year) and untreated arrhythmias (0.12 vs. 0.23/patient-year).
- The enhanced SMART Pass version demonstrated a lower deactivation rate (0.04 vs. 0.14/patient-year) and reduced treated and untreated arrhythmias.
Conclusions:
- S-ICD devices are typically programmed to detect high ventricular rates to minimize inappropriate shocks.
- SMART Pass activation is linked to reduced rates of detected and treated arrhythmias.
- The enhanced SMART Pass version appears to decrease deactivation rates and further reduce treated arrhythmias.
Keywords:
implantable defibrillatorinappropriate shockprogramingsubcutaneoussudden cardiac deathventricular arrhythmia