Cre8 Drug Eluting Stent Performance in Daily Cardiology Practice

Umit Yasar Sinan1, Ebru Serin1, Bengisu Keskin-Meric1

  • 1Department of Cardiology, İstanbul University Cardiology Institute, 34096 Istanbul, Turkey.

Insights

The Cre8 polymer-free amphilimus-eluting stent (PF-AES) demonstrated favorable clinical efficacy and safety in real-world cardiology practice. This study confirms its performance is comparable to newer drug-eluting stents (DES).

Area of Science:

  • Cardiology
  • Interventional Cardiology
  • Biomaterials Science

Background:

  • Drug-eluting stents (DES) are standard for percutaneous coronary intervention (PCI).
  • Key DES performance factors include stent design, alloy, polymer coating biocompatibility, and drug properties.
  • Cre8 is a polymer-free amphilimus-eluting stent (PF-AES).

Purpose of the Study:

  • To evaluate the clinical efficacy and safety of Cre8 PF-AES in routine cardiology practice.
  • To compare Cre8 stent performance against existing drug-eluting stents.

Main Methods:

  • Retrospective analysis of patients treated with Cre8 DES between December 2015 and 2016.
  • Inclusion criteria: chronic or acute coronary syndrome patients undergoing PCI with Cre8 DES.
  • Data collected on patient demographics, lesion characteristics, procedural success, and clinical outcomes.

Main Results:

  • 808 lesions in 664 patients were analyzed; mean follow-up was 487 days.
  • Procedural success rate was 97.3%.
  • Target lesion failure (TLF) occurred in 6.4% of lesions; cardiac death in 3.0%; target vessel revascularization (TVR) in 4.4%.

Conclusions:

  • Cre8 stents show clinical efficacy and safety in real-world settings.
  • Outcomes are comparable to latest-generation DES in randomized trials.
  • Diabetes and ejection fraction were identified as predictors of mortality and TLF.
Abstract