Effect of scan duration on CT perfusion values in metastases from renal cell carcinoma
Chaan S Ng1, Adam G Chandler2, Yanwen Chen3
1Department of Radiology, Unit 1473, University of Texas M.D. Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, TX 77030, USA.
Objective:
CT perfusion (CTp) values are affected by CT scan acquisition duration (tacq); their reproducibility is adversely affected by uncertainty in their measurement. The objectives were to assess the effects of tacq on CTp parameter values in metastases from renal cell carcinoma (mRCC) in thoracic and abdominal locations.
Materials And Methods:
131 CTp evaluations in 53 patients with mRCC were retrospectively analyzed by distributed parameter modeling to yield tissue blood flow (BF), blood volume (BV), mean transit time (MTT), permeability (PS), and also hepatic arterial perfusion (HAP) and hepatic arterial fraction (HAF) for liver metastases and normal liver, with tacq from 25 to 590 s. Penalized piecewise polynomial regression (SPLINE) characterized functional relationships between CTp parameters and acquisition duration, tacq. Evidence for time-invariance was evaluated for each parameter at multiple time points by conducting inference on the fitted derivative to assess its proximity to zero as a function of acquisition time. Equivalence testing was implemented with three levels of confidence (low (20%), moderate (70%), high (95%)).
Results:
Systematic and non-systematic variability was observed for CTp parameter values with limited tacq. All parameters in all locations approached increasing stability with increasing tacq. PS, HAP and HAF required longer acquisition times than BF, BV and MTT to attain comparable levels of stability. Stabilization tended to require longer acquisition in liver than other tissues. tacq=380 s was required to obtain at least moderate level of confidence for all parameters and organs.
Conclusion:
Increasing tacq yields increasingly more stable CT perfusion parameters, and thereby better reproducibility.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
09:00Spatial Measurements of Perfusion, Interstitial Fluid Pressure and Liposomes Accumulation in Solid Tumors
Published on: August 18, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET