Consequences of Pausing Heated Humidification During Invasive Ventilation
François Lellouche1, Pierre-Alexandre Bouchard2
1Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Québec, Canada. francois.lellouche@criucpq.ulaval.ca.
Background:
During invasive mechanical ventilation, where medical gases are very dry and the upper airway is bypassed, appropriate gas conditioning and humidification are mandatory at all times. Results of in vitro studies suggest that dry gases may improve lung deposition during nebulization, but this has not been confirmed through in vivo studies. The objective of this study was to measure gas humidity under multiple conditions to better describe gas hygrometry when heated humidifiers are turned off.
Methods:
We measured, on a bench, the hygrometry of different gases at steady state: medical gases, at the Y-piece without humidifier, with the humidifier switched off, and with humidifier switched on. We measured gas humidity every 10-60 s during dynamic conditions after switching off the heated humidifier and after switching on the heated humidifier. Hygrometry was measured by using the psychrometric method with at least 3 measurements for each tested condition.
Results:
We performed 287 psychrometric measurements in different situations. The mean ± SD gas absolute humidity at steady state during different conditions were the following: 1.6 ± 0.2 mg H2O/L for the medical gases, 4.5 ± 0.9 mg H2O/L at the Y-piece without humidifier, 9.1 ± 0.3 mg H2O/L at the Y-piece with heated humidifier turned off, and 34.2 ± 2.2 mg H2O/L at the Y-piece with the heated humidifier turned on. During the dynamic evaluation, after turning off the humidifier, humidity was < 30 mg H2O/L after a few minutes, attained 15 mg H2O/L after 15 min, and was below 10 mg H2O/L after 1 h but never reached the level of dry medical gases. After turning on the heated humidifier, the gas hygrometry reached 30 mg H2O/L after 5 min.
Conclusions:
When heated humidifiers are turned off, gas humidity levels are very low but not as low as medical gases. The clinical impact of repeated shutdowns is unknown. As recommended, heated humidifiers should never be turned off during nebulization.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Hyperpnea and Hyperventilation
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....