Validating the accuracy of the Omron HBP-M4500 blood pressure measuring device according to the ISO81060-2:2018+ Amd
Hakuo Takahashi1, Shingo Yamashita2,1, Nobuki Yakura2,1
1Department of Cardiology, Biwako Yoikuin Hospital, Shiga, Japan.
Background:
Blood pressure (BP) naturally undergoes fluctuations and variations, particularly during anesthesia administration during surgery, hemodialysis, upper and lower gastrointestinal endoscopy, exercise testing, arterial and venous catheterization, and rehabilitation. These changes in BP may lead to life-threatening events.
Objectives:
The performance of the Omron HBP-M4500 device in monitoring blood pressure (BP) in the upper arm was validated according to the International Organization for Standardization (ISO) 81060-2:2018+amendment (Amd) 1:2020 protocol.
Methods:
The device was used to assess 113 participants in the inflation mode, and 107 participants in the deflation mode. All the patients fulfilled the inclusion criteria, including the arm circumference range and systolic and diastolic BP levels, outlined in the protocol. Data validation and analysis were performed according to the manufacturer's instructions.
Results:
In criterion 1, the mean ± standard deviation (SD) values of the differences between the test device and reference BP were -0.6 ± 5.80/2.8 ± 6.78 mmHg (systolic/diastolic) and -1.0 ± 5.35/3.2 ± 6.52 mmHg for the inflation and deflation modes, respectively. These data fulfilled the ISO81060-2:2018+Amd1:2020 requirements of ⩽ 5 ±⩽ 8 mmHg. In criterion 2, the differences were -0.6 ± 4.44/2.8 ± 6.26 and -1.0 ± 3.84/3.2 ± 6.09 mmHg for the inflation and deflation modes, respectively, fulfilling criterion 2 with SD values of ⩽ 6.91 and ⩽ 6.87 for systolic BP and ⩽ 6.34 and ⩽ 6.14 for diastolic BP in the inflation and deflation modes, respectively. These two criteria were fulfilled in both studies.
Conclusion:
The Omron HBP-M4500 device, either in inflation or deflation mode, fulfilled the criteria outlined in the ISO protocol. Therefore, this device is valuable for BP measurement in clinical and hospital settings.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)