The effect of reconstruction algorithms on semi-quantitative measurements in 18F-FDG PET/CT imaging
Filiz Özülker1, Gündüzalp Buğrahan Babacan, Safiya Cengiz
1University of Health Sciences, Prof. Dr. CemilTaşcıoğlu Hospital, Department of Nuclear Medicine, İstanbul, Turkey. fozulker@gmail.com.
Objective:
This study was carried out to understand whether Q.Clear and ordered subset expectation maximization (OSEM), reconstruction algorithms used in fluorine-18-fluorodeoxyglucose (18F-FDG) positron emission tomography/computed tomography (PET/CT) applications, and parameters such as time of flight (TOF) and point spread function (PSF) cause different results in semi-quantitative measurements.
Subjects And Methods:
Raw PET data of 264 patients who were referred to 18F-FDG PET/CT imaging with the purpose of evaluation of known or suspicious malignant disease were reconstructed separately with Q.Clear (GE Healthcare), a BPL, an OSEM algorithm, PSF (SharpIR®) and TOF (VUE Point FX®) methods. Each patient's liver, mediastinal blood pool, metabolic tumor volume (MTV), total lesion glycolysis (TLG), and standardized uptake values (SUV) (SUVmax, SUVmean, and SUVpeak) of a total of 264 lesions selected from the patients were performed.
Results:
β350+ToF yielded higher measurement results than all other variables for all of the lesion SUVmax, lesion SUVmean, L/AP SUVmax, and L/AP SUVmean parameters. OSEM+ToF and OSEM+TOF+PSF algorithms yielded higher mean and median SUVmax values for the reference structures (liver and mediastinum) and for lesions SUVmax and SUVmean values were statistically significantly lower than the β350+ToF method. The method with the lowest mean value for the L/Liver SUVmax variable was OSEM+ToF 4iter16ss (mean=1.76), while the method with the highest mean value was β350+ToF (mean=2.26). β350+ToF was the reconstruction method with the highest ratios for L/AP SUVmax and SUVmean for both lesions below and above 1 cm. β350+ToF algorithm had also statistically significantly higher results for these variables compared to all other parameters in malignant lesions.
Conclusion:
When comparing 18F-FDG PET/CT images, the use of different reconstruction algorithms may lead to misleading results, especially in the evaluation of response to treatment of malignancies.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET