Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Simultaneous whole-liver water T 1 and T 2 mapping with isotropic resolution during free-breathing
Jonathan Stelter1, Kilian Weiss2, Lisa Steinhelfer1
1Institute of Diagnostic and Interventional Radiology, School of Medicine and Health, Technical University of Munich, Munich, Germany.
Purpose:
To develop and validate a data acquisition scheme combined with a motion-resolved reconstruction and dictionary-matching-based parameter estimation to enable free-breathing isotropic resolution self-navigated whole-liver simultaneous water-specific ( ) and ( ) mapping for the characterization of diffuse and oncological liver diseases.
Methods:
The proposed data acquisition consists of a magnetization preparation pulse and a two-echo gradient echo readout with a radial stack-of-stars trajectory, repeated with different preparations to achieve different and contrasts in a fixed acquisition time of 6 min. Regularized reconstruction was performed using self-navigation to account for motion during the free-breathing acquisition, followed by water-fat separation. Bloch simulations of the sequence were applied to optimize the sequence timing for insensitivity at 3 T, to correct for relaxation-induced blurring, and to map and using a dictionary. The proposed method was validated on a water-fat phantom with varying relaxation properties and in 10 volunteers against imaging and spectroscopy reference values. The performance and robustness of the proposed method were evaluated in five patients with abdominal pathologies.
Results:
Simulations demonstrate good insensitivity of the proposed method in measuring and values. The proposed method produces co-registered and maps with a good agreement with reference methods (phantom: ; ). The proposed and mapping exhibits good repeatability and can be robustly performed in patients with pathologies.
Conclusions:
The proposed method allows whole-liver and quantification with high accuracy at isotropic resolution in a fixed acquisition time during free-breathing.