MyotonPRO as a new valid tool for measuring cervical muscle tension. A reliability study
Sol Ferran de la Cierva1, David Terrasa1, Octavio Garaycochea1
1Otorhinolaryngology Department, Clínica Universidad de Navarra. Pamplona, Spain.
Logopedics, Phoniatrics, Vocology
|August 9, 2024
Summary
This study shows that using a myotonometer to measure neck and face muscle properties is reliable for voice disorder diagnosis. Further research can confirm its use in objective assessment and treatment monitoring.
Area of Science:
- Biomechanics
- Speech-Language Pathology
- Medical Device Technology
Background:
- Muscle tension dysphonia is a voice disorder often diagnosed subjectively.
- Objective biomechanical measurements of relevant muscles could aid diagnosis and treatment evaluation.
- Non-invasive methods are desirable for assessing cervicofacial muscle function.
Purpose of the Study:
- To assess the reliability of a handheld myotonometer for measuring biomechanical properties of cervicofacial muscles.
- To evaluate the potential of this method for diagnosing voice disorders, particularly muscle tension dysphonia.
- To determine inter- and intra-rater reliability of the myotonometer measurements.
Main Methods:
- A prospective pilot study involving 10 subjects.
- Measurements of frequency, stiffness, elasticity, relaxation time, and creep using a MyotonPRO handheld myotonometer.
- Inclusion of masseter, sternocleidomastoid, orbicularis oris, semispinalis capitis, suprahyoid, infrahyoid, and trapezius muscles.
- Inter- and intra-rater reliability assessed by two evaluators.
Main Results:
- Good to excellent inter-rater reliability was found for masseter, sternocleidomastoid, trapezius, and suprahyoid muscles.
- Good to excellent intra-rater reliability was observed for masseter, sternocleidomastoid, semispinalis capitis, and suprahyoid muscles.
- Lower reliability (ICC) was noted for stiffness and creep properties across several muscles.
Conclusions:
- Myotonometer measurement of mechanical properties in cervical and orofacial muscles is a reliable and reproducible technique.
- Further research is needed to link muscle properties to voice disorders and validate the tool for objective diagnosis and treatment monitoring.