A New Versatile Jig for the Calibration and Validation of Force Metrics with Instrumented Paddles in Sprint Kayaking
Hans Rosdahl1, David Aitken1, Mark Osborne2
1The Swedish School of Sport and Health Sciences (GIH), SE-114 86 Stockholm, Sweden.
Abstract:
The interest in using new technologies to obtain recordings of on-water kinetic variables for assessing the performance of elite sprint kayakers has increased over the last decades but systematic approaches are warranted to ensure the validity and reliability of these measures. This study has an innovative approach, and the aim was to develop a new versatile jig including reference force sensors for both the calibration and validation of mutual static and dynamic stroke forces as measured with instrumented paddles at the high force levels used in elite sprint kayaking.
Methods:
A jig was constructed using a modified gym weight stack and a frame consisting of aluminum profiles permitting a fastening of custom-made kayak paddle shaft and blade support devices with certified force transducers combined with a data acquisition system to record blade and hand forces during static (constant load) and dynamic conditions (by paddle stroke simulation). A linear motion path incorporating a ball-bearing equipped carriage with sensors for the measurement of vertical distance and horizontal displacement was attached to the frame for recordings of various position measures on the paddle. The jig design with all components is extensively described to permit replication. The procedures for assessing the accuracy of the jig force instrumentation are reported, and with one brand of instrumented paddle used as an example, methods are described for force calibration and validation during static and dynamic conditions.
Results:
The results illustrate that the measured force with the jig instrumentation was similar to the applied force, calculated from the applied accurate mass (within a -1.4 to 1.8% difference) and similar to the force as calculated from the applied mass with the weight stack (within a -0.57 to 1.16% difference). The jig was suitable for the calibration and validation of forces in a range relevant for elite sprint kayaking under both static and dynamic conditions. During static conditions with a force direction equal to the calibration conditions and a force range from 98 to 590 N, all values for the instrumented paddle were within a -3.4 to 3.0% difference from the jig sensor values and 28 of 36 values were within ±2%. During dynamic conditions with paddle stroke simulations at 60 and 100 strokes/min and a target peak force of 400 N, the common force variables as measured by the instrumented paddle were not significantly different from the same measures by the jig (values at 100 strokes/min: peak force; 406.9 ± 18.4 vs. 401.9 ± 17.2 N, mean force; 212.8 ± 15.4 vs. 212.0 ± 14.4 N, time to peak force; 0.17 ± 0.02 vs. 0.18 ± 0.02 s, force impulse; 90.8 ± 11.2 vs. 90.5 ± 10.8 Ns, impulse duration; 0.43 ± 0.03 vs. 0.43 ± 0.03 s).
Conclusion:
A novel jig with several new functions is presented that enables the calibration and validation of force measurements with instrumented paddles by providing standardized conditions for calibration and force validation during both static and dynamic conditions in a force range relevant to elite sprint kayaking.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Three-Dimensional Force System
Constant Pressure Calorimetry
Two-Dimensional Force System
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...