Automated Method for Growing Rod Length Measurement on Ultrasound Images in Children With Early Onset Scoliosis
Mohammad Humayun Kabir1, Marek Reformat1, Sarah Southon Hryniuk2
1Department of Electrical and Computer Engineering, University of Alberta, Edmonton, AB, Canada.
Objective:
To develop and validate machine learning algorithms to automatically extract the rod length of the magnetically controlled growing rod from ultrasound images (US) in a pilot study.
Methods:
Two machine-learning (ML) models, called the "Boundary model" and "Rod model," were developed to identify specific rod segments on ultrasound images. The models were developed utilizing Mask Regional Convolutional Neural Networks (Mask RCNN). Ninety US images were acquired from 23 participants who had early onset scoliosis (EOS) surgeries; among those, 70 were used for model development, including training and validation, and 20 were used for testing by comparing the AI-based vs. manual measurements.
Results:
The average precision (AP) of the ML models was 88.5% and 60.2%, respectively. The inter-method correlation coefficient (ICC) was 0.98, and the mean absolute difference ± standard deviation (MAD ± SD) between AI and manual measurements was 0.86 ± 1.0 mm. The Bland-Altman analysis showed no bias, and 90% of the data were within the 95% confidence interval. The automated method was reliable, accurate, and fast. Measurements were displayed in 4.6 seconds after the US image was inputted.
Conclusion:
This was the first AI-based method to measure the MCGR rod length on US images automatically.