Foam-based Left Atrial Appendage Closure (CLAAS) Device: Evaluation in a Chronic Canine Model.
Aaron V Kaplan1,2, Carol Devellian2, Jamie H Kim3
1Heart & Vascular Center, Dartmouth Hitchcock Medical Center, Lebanon, New Hampshire.
Summary
The Conformal Left Atrial Appendage System (CLAAS) implant effectively seals the left atrial appendage in canines for 150 days. This novel device shows no signs of local or systemic toxicity, indicating its safety and efficacy for stroke prevention.
Area of Science:
- Cardiovascular Devices
- Biomaterials Science
- Medical Device Engineering
Background:
- The Conformal Left Atrial Appendage System (CLAAS) is a transcatheter implant designed for left atrial appendage occlusion.
- It aims to prevent stroke in patients with nonvalvular atrial fibrillation.
- The implant features a novel polyurethane polycarbonate-urea foam matrix for enhanced performance.
Purpose of the Study:
- To characterize the thrombogenicity, sealing efficacy, and implant integrity of the CLAAS device.
- To evaluate local tissue response and systemic toxicity in a canine model.
- To assess the long-term performance of the CLAAS implant up to 150 days.
Main Methods:
- The CLAAS implant was evaluated in a chronic canine model (n=8) using echocardiographic and fluoroscopic guidance.
- Postprocedural echocardiographic assessments were conducted at multiple time points up to 150 days.
- Necropsy, including gross, x-ray, and histologic examination, was performed at 90 and 150 days.
Main Results:
- Successful implantation was achieved in all animals, with complete sealing of the left atrial appendages observed on gross examination.
- Minor leaks (<5 mm) were noted in two animals, one at implantation and one at day 90.
- No device-related thrombus, pericardial effusion, distal organ emboli, or histologic evidence of toxicity was observed.
Conclusions:
- The CLAAS implant demonstrates successful deployment and effective sealing of the left atrial appendage in a canine model for up to 150 days.
- The device maintained structural integrity and showed good tissue integration.
- No local or systemic toxicity was observed, supporting its potential for clinical use in stroke prevention.