Virtual non-contrast images in photon-counting computed tomography: impact of different contrast phases
Eva Laurin Gadsbøll1, Marie-Louise Aurumskjöld2,3, Fredrik Holmquist1,4
1Department of Imaging and Functional Medicine, Skåne University Hospital, Lund, Sweden.
Background:
Photon-counting computed tomography (PCCT) enables new ways of image reconstruction, e.g. material decomposition and creation of virtual non-contrast (VNC) series with higher resolution and lower radiation dose than standard computed tomography (CT). Clinical experiences of this are limited.
Purpose:
To compare true non-contrast (TNC) series with VNC series derived from non-enhanced (VNCu), arterial phase (VNCa) and portal venous phase (VNCv) in clinically approved PCCT.
Material And Methods:
A total of 45 clinical, tri-phasic abdominal CT scans from the PCCT Naetom Alpha, between February 2022 and November 2022, were retrospectively assessed. Placing a region of interest in six different locations in each VNC series - right liver parenchyma, left liver parenchyma, spleen, aorta, erector spinae muscle, and in the subcutaneous fat - absolute Hounsfield values (HU) and standard deviations (SD) were collected. Differences in HU (ΔHU) were compared and statistically analyzed.
Results:
Statistically significant differences between VNC and TNC were seen in all measurements, with the largest difference in the subcutaneous fat and the smallest difference in the erector spinae muscle. Only small differences were seen between VNCa and VNCv, where the largest differences were seen in the left and right liver lobes.
Conclusion:
VNC images from the first-generation clinically approved PCCT showed a significant difference between VNC and TNC images. The differences vary with the type of tissue. Only small differences were seen depending from which contrast phase the VNC was derived.
More Related Videos
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging