Related Experiment Video
Updated: Jun 16, 2025
![Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F58491.jpg&w=3840&q=50)
Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
Published on: August 8, 2019
An accelerated PETALUTE MRI sequence for in vivo quantification of sodium content in human articular cartilage at 3T
Cameron X Villarreal1, Xin Shen1,2, Ahmad A Alhulail3
1Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, 47907, USA.
Objective:
In this work, we evaluate the sodium magnetic resonance imaging (MRI) capabilities of a three-dimensional (3D) dual-echo ultrashort echo time (UTE) sequence with a novel rosette petal trajectory (PETALUTE), in comparison to the 3D density-adapted (DA) radial spokes UTE sequence in human articular cartilage in the knee.
Materials And Methods:
We scanned five healthy subjects using a 3D dual-echo PETALUTE acquisition and two comparable implementations of 3D DA-radial spokes acquisitions, one matching the number of k-space projections (Radial - Matched Spokes) and the other matching the total number of samples (Radial - Matched Samples) acquired in k-space.
Results:
The PETALUTE acquisition enabled equivalent sodium quantification in articular cartilage volumes of interest (168.8 ± 29.9 mM, mean ± standard deviation) to those derived from the 3D radial acquisitions (171.62 ± 28.7 mM and 149.8 ± 22.2 mM, respectively). We achieved a 41% shorter scan time of 2:06 for 3D PETALUTE, compared to 3:36 for 3D radial acquisitions. We also evaluated the feasibility of further acceleration of the PETALUTE sequence through retrospective compressed sensing with 2 × and 4 × acceleration of the first echo and showed structural similarity of 0.89 ± 0.03 and 0.87 ± 0.03 when compared to non-retrospectively accelerated reconstruction.
Conclusion:
We demonstrate improved scan time with equivalent performance using a 3D dual-echo PETALUTE sequence compared to the 3D DA-radial sequence for sodium MRI of articular cartilage.
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

