Advanced Lung Imaging with Photon-counting Detectors: Insights from Thermoluminescence Dosimetry
Henner Huflage1, Carsten Hackenbroch2, Simone Schüle2
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080 Würzburg, Germany.
Rationale And Objectives:
This study investigates the dose burden of photon-counting detector (PCD) lung CT with ultra-high-resolution (UHR) and standard mode using organ-based tube current modulation (OBTCM).
Materials And Methods:
An anthropomorphic Alderson-Rando phantom was scanned in UHR and standard mode with and without OBTCM on three dose levels (IQ 5, 20, 50). Effective radiation dose was determined by thermoluminescent dosimetry in 13 measurement sites and compared with the calculated effective dose derived from the dose-length product. Image quality was evaluated subjectively by six radiologists using an equidistant 7-point scale and objectively by means of modulation transfer function analysis.
Results:
Measured effective radiation exposure was lower in UHR and OBTCM studies than in standard mode (IQ 5: 0.34-0.36, IQ 20: 1.57-1.70, IQ 50: 3.76-3.99 mSv). Compared with the calculated effective dose, the radiation exposure measured with thermoluminescence dosimetry was 131-170% higher. Noise in UHR mode was rated lower than in standard (all p ≤ 0.042) and OBTCM images (all p ≤ 0.028) for all dose levels, while image sharpness was deemed highest for UHR protocols (all p ≤ 0.042). The use of OBTCM had no significant effect on either dimension of subjective image quality (all p ≥ 0.999). Modulation transfer function analysis confirmed the highest spatial frequency in UHR datasets (all p ≤ 0.016).
Conclusion:
In PCD-CT of the lung, full field-of-view UHR imaging entails no dose disadvantage over standard mode despite superior image quality. OBTCM possesses moderate dose saving potential. Thermoluminescence dosimetry yielded considerably higher effective doses than those calculated from dose-length products.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Photoluminescence: Applications
Flame Photometry: Overview
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...