Biomechanical Assessment of Liver Integrity: Prospective Evaluation of Mechanical Versus Acoustic MR Elastography
Vitali Koch1, Jennifer Gotta1, Victoria Chernyak2
1Department of Diagnostic and Interventional Radiology, Goethe University Frankfurt, University Hospital, Frankfurt am Main, Germany.
Background:
Magnetic resonance elastography (MRE) can quantify tissue biomechanics noninvasively, including pathological hepatic states like metabolic dysfunction-associated steatohepatitis.
Purpose:
To compare the performance of 2D/3D-MRE using the gravitational (GT) transducer concept with the current commercial acoustic (AC) solution utilizing a 2D-MRE approach. Additionally, quality index markers (QIs) were proposed to identify image pixels with sufficient quality for reliably estimating tissue biomechanics.
Study Type:
Prospective.
Population:
One hundred seventy participants with suspected or confirmed liver disease (median age, 57 years [interquartile range (IQR), 46-65]; 66 females), and 11 healthy volunteers (median age, 31 years [IQR, 27-34]; 5 females).
Field Strength/Sequence:
Participants were scanned twice at 1.5 T and 60 Hz vibration frequency: first, using AC-MRE (2D-MRE, spin-echo EPI sequence, 11 seconds breath-hold), and second, using GT-MRE (2D- and 3D-MRE, gradient-echo sequence, 14 seconds breath-hold).
Assessment:
Image analysis was performed by four independent radiologists and one biomedical engineer. Additionally, superimposed analytic plane shear waves of known wavelength and attenuation at fixed shear modulus were used to propose pertinent QIs.
Statistical Tests:
Spearman's correlation coefficient (r) was applied to assess the correlation between modalities. Interreader reproducibility was evaluated using Bland-Altman bias and reproducibility coefficients. P-values <0.05 were considered statistically significant.
Results:
Liver stiffness quantified via GT-2D/3D correlated well with AC-2D (r ≥ 0.89 [95% CI: 0.85-0.92]) and histopathological grading (r ≥ 0.84 [95% CI: 0.72-0.91]), demonstrating excellent agreement in Bland-Altman plots and between readers (κ ≥ 0.86 [95% CI: 0.81-0.91]). However, GT-2D showed a bias in overestimating stiffness compared to GT-3D. Proposed QIs enabled the identification of pixels deviating beyond 10% from true stiffness based on a combination of total wave amplitude, temporal sinusoidal nonlinearity, and wave signal-to-noise ratio for GT-3D.
Conclusion:
GT-MRE represents an alternative to AC-MRE for noninvasive liver tissue characterization. Both GT-2D and 3D approaches correlated strongly with the established commercial approach, offering advanced capabilities in abdominal imaging compared to AC-MRE.
Evidence Level:
1 TECHNICAL EFFICACY: Stage 2.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection....
Assessment of the Abdomen III: Palpation