Related Experiment Video
Updated: Aug 10, 2026

Microfluidic Platform with Multiplexed Electronic Detection for Spatial Tracking of Particles
Published on: March 13, 2017
An Electrochemical CMOS Biosensor Array Using Phase-Only Modulation With 0.035% Phase Error and In-Pixel Averaging
Abstract:
This paper presents a 16 × 20 CMOS biosensor array based on electrochemical impedance spectroscopy (EIS), a highly sensitive label-free technique for rapid disease detection at the point-of-care. This high-density system implements polar-mode detection with phase-only EIS measurement over a 5 kHz - 1 MHz frequency range. The design features predominantly digital readout circuitry, ensuring scalability with technology, along with a load-compensated transimpedance amplifier, all within a 140 × 140 µm2 pixel. The architecture enables in-pixel digitization and accumulation, which increases the SNR by 10 dB for each 10× increase in readout time. Implemented in a 180 nm CMOS process, the 3 × 4 mm2 chip achieves state-of-the-art performance with an rms phase error of 0.035% at 100 kHz through a duty-cycle insensitive phase detector and one of the smallest per pixel areas with in-pixel quantization.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
10:05In Vitro Multiparametric Cellular Analysis by Micro Organic Charge-modulated Field-effect Transistor Arrays
Published on: September 20, 2021
Related Concept Videos
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Microbial Biosensors