Comparative Study Between Blood Glucose Meters in a Maternity Center
Rihab Ayadi1, Monia Tlig1, Imène Ben Jdida2
1Faculty of Pharmacy of Monastir, Monastir, Tunisia.
Background:
Monitoring of diabetes by measuring capillary blood glucose using a glucometer.
Objectives:
To compare the three most used glucose meters in diabetic patients in a maternity and neonatal center in terms of repeatability, accuracy, and stability.
Methodology:
A comparative study on 100 diabetic patients admitted to the various departments of a maternity and neonatal center. For each patient, a capillary measurement was made using each of the three glucometers to be tested (Accucheck®, On Call® and Bionime®) as well as a blood glucose on venous blood, performed in the laboratory using the Siemens X brand plus® PLC (reference method). The same sample was used to carry out all measurements.
Results And Conclusions:
The Accucheck® brand reader and the On Call® brand thus, show a better correlation for open and closed bottles of test strips'storage, compared with the Bionime® brand reader. We found that the Accucheck® and On Call® readers are comparable in terms of reliability of results and are better than the Bionime® brand reader. We have found that, first using closed vials give more accurate results for glycemia readers and also, On Call® and Accucheck® brand readers are more reliable and accurate than the Bionime® brand reader.
More Related Videos
Related Concept Videos
Diabetes Mellitus: Type 2 and Gestational
Pathophysiology of Diabetes
Type 1 diabetes is characterized by autoimmune-mediated destruction of pancreatic β cells, with environmental factors potentially triggering this process in genetically susceptible individuals. Despite many not having a family history, certain genes increase susceptibility,...