Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Prospective Randomized Trial Comparing 2 Devices for Deep Inspiration Breath Hold Management in Breast Radiation
Hilary L Byrne1, Elisabeth Steiner1,2, Jeremy Booth3,4
1Image X Institute, Sydney School of Health Sciences, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
Purpose:
The Breast Radiotherapy Audio Visual Enhancement for sparing the Heart (BRAVEHeart) trial prospectively randomized patients with left-sided breast cancer to 1 of 2 deep inspiration breath hold biofeedback devices: a novel chest surface tracking system and an abdominal block tracking system. The primary hypothesis was that the accuracy of chest tracking would be higher than that of abdominal tracking as the chest is a more direct surrogate of the breast target.
Methods And Materials:
Patients with left-sided breast cancer were treated in deep inspiration breath hold with intensity modulated radiation therapy delivery. Patients were randomized to either the novel chest surface system or abdominal block system for active management of breath hold with visual feedback. On both trial arms, the unallocated system was monitored passively. A total of 239,296 cine electronic portal imaging device images were analyzed retrospectively to extract the chest wall position. Treatment accuracy was quantified as the deviation of the internal chest wall during treatment relative to the planned position from the digitally reconstructed radiograph. The correlation between motion of the external surrogate and internal chest wall was calculated per-breath hold. Ease of use was assessed with questionnaires for both radiation therapists and patients and appointment length recorded.
Results:
Data from 26 participants were available for analysis. No difference was found in delivered treatment accuracy between arms. Across all patients and fractions, the median correlation between internal chest wall movement and external surrogate was 0.69 for the chest surface and 0.17 for the abdominal block. Patients found it easy to follow visual feedback from both systems. No difference was found in appointment length between arms.
Conclusions:
No statistical evidence was found for superior treatment accuracy, satisfaction, or appointment length for the novel chest surface tracking device compared with the abdominal block system. During deep inspiration breath hold, the median per-breath hold correlation of internal chest wall movement to the motion of the chest surface was higher than the median correlation of the abdominal block to the chest surface.
More Related Videos
08:34Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Other Factors Affecting Respiration Centers
However, the ability to hold one's breath voluntarily is not limitless. When the CO2 concentration in the blood reaches a critical...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...