Multi-site phantomless bone mineral density from clinical quantitative computed tomography in males
Zachary A Haverfield1, Amanda M Agnew1, Kathryn Loftis2
1Injury Biomechanics Research Center, The Ohio State University, Columbus, Ohio 43210, United States.
Phantomless methods for calculating volumetric bone mineral density (vBMD) using QCT are accurate and comparable to phantom-based methods. Regression-based phantomless (RPL) calibration shows promise for opportunistic QCT and assessing bone quality.
Area of Science:
- Radiology
- Orthopedics
- Biomedical Engineering
Background:
- Quantitative CT (QCT) is standard for volumetric bone mineral density (vBMD) assessment.
- Traditional QCT requires phantom rods, which are not always available.
- Phantomless methods, including internal-reference phantomless (IPL) and Hounsfield unit (HU) measurements, offer alternatives.
Purpose of the Study:
- To evaluate the accuracy of phantomless QCT methods (IPL and regression-based phantomless (RPL) using HU) compared to phantom-based (PB) methods.
- To assess the performance of these methods across multiple skeletal sites.
- To validate HU as a proxy for vBMD calculation.
Main Methods:
- QCT scans from 100 male post-mortem human subjects (PMHS) were analyzed.
- vBMD was calculated using site-specific PB calibration.
- IPL and RPL vBMD were determined using a development sample (50/100 PMHS) for calibration and validated on the remaining subjects.
Main Results:
- PB and phantomless vBMD (IPL/RPL) showed no significant differences (p > .05).
- Univariate regressions between PB and phantomless vBMD were significant (p < .05), except for IPL Rad-30 (p = .078).
- RPL vBMD demonstrated strong correlation with PB vBMD (R² = 99.17%) with smaller differences compared to IPL vBMD (R² = 97.99%).
Conclusions:
- Phantomless QCT methods, particularly RPL, are accurate and comparable to PB methods for vBMD assessment.
- Site-specific calibration is recommended for phantomless methods to enhance accuracy.
- RPL calibration improves the utility of opportunistic QCT for bone quality and fracture risk assessment.