MARPE expander activation load with different configurations of extender arms heights: in-vitro evaluation
Staell Ribeiro de Faria1, Túlio Rodrigues de Andrade1, Cristiane Barros André2
1Faculdade São Leopoldo Mandic, Dentistry Department (Campinas/SP, Brazil).
Objective:
Evaluate the load mini-implants exert on the artificial bone when expanding the MARPE EX in three different extension arm configurations.
Methods:
A device simulating the human palate was fabricated and attached to a universal testing machine, for conducting tests with different MARPE expanders (n=5): non-adjustable/control (MARPE SL, Peclab) or with low, intermediate, and high extender arms (MARPE EX, Peclab). The expanders were manually activated until failure of the device occurred, and maximum load values were recorded. Load averages were also calculated for every five activations until the twentieth activation.
Results:
The generalized linear mixed model for repeated measures over time showed that there was significant increase in load with activations for all expanders (p=0.0004). Up to the twentieth activation, the expander with low extender arms presented higher load than the others, while the expander with high extender arms showed lower load values (p<0.05). There was no significant difference among expanders regarding the number of activations (p=0.0586), although there was a trend towards fewer activations until fracture for the control expander. It was observed that the higher the configuration, the lower the force the mini-implants delivered to the bone. The control expander provided a force magnitude similar to that of the adjustable expander when positioned at the intermediate height.
Conclusions:
The activation load of MARPE expanders is influenced by the type of presentation of the extensor arms, with higher configurations resulting in lower force delivered by the mini-implants to the bone.
More Related Videos
07:03Author Spotlight: Bridging the Gap Between In Vivo and Ex Vivo Studies with the "Avatar" Technique to Advance Muscle Mechanics Research
Published on: August 18, 2023
07:16Author Spotlight: Development of a Novel Finite Element Analysis Model for Improved Orthognathic Surgical Techniques
Published on: October 20, 2023
Related Concept Videos
Machines: Problem Solving II
Cable Subjected to a Distributed Load
Applications of Stress
The...
Design Consideration
The factor of safety is another key aspect...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...