A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
A Raman measurement scheme for accurate compositional analysis of dark-colored sample by minimizing laser-induced
Sanghoon Cho1, Haeseong Jeong1, Yoonjeong Lee2
1Department of Chemistry and Research Institute for Convergence of Basic Science, Hanyang University, Seoul, 04763, Republic of Korea.
Abstract:
An effective and rapid Raman measurement scheme to determine Fe3O4 concentration in sintered ores was explored. Because sintered ores are brownish-black materials that easily absorb laser photons, accurate quantitative analysis requires obtaining an Fe3O4 peak with a high signal-to-ratio by reducing the possibility of local sample heating and degradation. For this purpose, a wide area coverage (WAC) Raman scheme with a laser-illumination diameter of 1 mm was adopted to decrease the laser power per area (LP/A) on each sample. The sintered ore sample was also wetted with water to reduce the chance of further heating by the laser. The combination of the WAC scheme and water-wetting allowed to increase the laser power during sample measurement, and the subsequent intensity (as well as the signal-to-noise ratio) of the Fe3O4 peak was elevated compared with both that measured by a Raman microscope yielding a higher LP/A and without water-wetting of the sample. In the Raman spectra of 93 real sintered ore samples measured using the proposed scheme, the ratio of Fe3O4 and Fe2O3 peak areas correlated closely (R2 = 0.94) with Fe3O4 concentration determined by titration. The demonstrated scheme is practical when Raman spectroscopy is employed for compositional analysis of dark and highly photon-absorbing samples.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...