Precise Determination of Cd Isotope Ratios at 3-10 ng Level by Thermal Ionization Mass Spectrometry Using a
Chao-Feng Li1,2, Zhu-Yin Chu1,2, Peng Peng1,2
1State Key Laboratory of Lithospheric Evolution, Institute of Geology and Geophysics, Chinese Academy of Sciences, Beijing 100029, China.
Abstract:
Thermal ionization mass spectrometry (TIMS) combined with the double spike technique has excellent analytical precision for Cd isotopic ratio analysis. However, because of the low ionization efficiency of Cd by TIMS, it is still not possible to obtain high precision Cd isotope ratios for small sample size (<100 ng) due to the lack of a highly sensitive emitter for Cd. A new loading method using molybdenum silicide (MoSi2) emitter has been developed for Cd isotope ratio measurements. This emitter produces a significant enhancement in the ionization efficiency of Cd and thus significantly reduces the required sample size to the 3-10 ng level. Analyses of δ114/110Cd for the NIST SRM 3108 using 108Cd-116Cd double spike method show excellent reproducibility (2 SD) that reaches ±0.032‰, ±0.042‰, and ±0.051‰ for 10, 5, and 3 ng of Cd, respectively. This method was further verified with a suite of geological reference materials. Replicate digestions and analyses (n = 8, 2 SD) of δ114/110Cd for NIST SRM 2711a, NOD A-1, and GBW08401 demonstrated good external reproducibility with results of 0.596 ± 0.024‰ for NIST SRM 2711a, 0.150 ± 0.036‰ for NOD A-1, and -0.665 ± 0.084‰ for GBW08401. These data clearly indicate that MoSi2 is an excellent alternative for traditional silica gel to Cd isotopic measurements, especially for samples with a low content of Cd.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Gas Chromatography: Types of Detectors-II
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...