Background:
The goal of this study was to develop and validate a UPLC-MS/MS method for simultaneous mea-surement of 13 AEDs, including carbamazepine, oxcarbazepine, lamotrigine, levetiracetam, topiramate, primidone, zonisamide, gabapentin, lacosamide, perampanel, pregabalin, rufinamide, and vigabatrin, in whole blood samples.
Methods:
A UPLC-MS/MS method for simultaneous determination of 13 AEDs in whole blood was developed, and validation was conducted for accuracy, precision, limit of quantification (LOQ), matrix effect, and stability. Our method was compared to two different hospitals using UPLC-MS/MS.
Results:
All AEDs exhibited linearity across the AMR (analytical measurement range), with R2 values ranging from 0.994 to 1.000. The imprecision and inaccuracy for low and high quality control (QC) levels were within an acceptable range, with the coefficient of variation (CV) < 15%. The LOQ was 0.62 µg/mL for carbamazepine, 1.61 µg/mL for oxcarbazepine, 1.30 µg/mL for lamotrigine, 13.20 µg/mL for levetiracetam, 1.26 µg/mL for topira-mate, 1.01 µg/mL for primidone, 1.59 µg/mL for zonisamide, 1.09 µg/mL for lacosamide, 1.61 µg/mL for gabapentin, 0.50 µg/mL for pregabalin, 0.07 ng/mL for perampanel, 3.00 µg/mL for rufinamide, and 2.06 µg/mL for vigabatrin. All AEDs demonstrated acceptable assay parameters for carryover, stability, and matrix effects. Moreover, the assay showed satisfactory results compared to two different hospitals with a bias of less than 15%.
Conclusions:
We successfully developed and validated a fast and robust UPLC-MS/MS method for routine therapeutic drug monitoring of thirteen antiepileptic drugs simultaneously.