On the correction of spiral trajectories on a preclinical MRI scanner with a high-performance gradient insert
Hannah Scholten1, Tobias Wech1,2, Sascha Köhler3
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany.
Abstract:
This study aimed to examine different trajectory correction methods for spiral imaging on a preclinical scanner with high-performance gradients with respect to image quality in a phantom and in vivo. The gold standard method of measuring the trajectories in a separate experiment is compared to an isotropic delay-correction, a correction using the gradient system transfer function (GSTF), and a combination of the two. Three different spiral trajectories, with 96, 16, and three interleaves, are considered. The best image quality is consistently achieved when determining the trajectory in a separate phantom measurement. However, especially for the spiral with 96 interleaves, the other correction methods lead to almost comparable results. Remaining imperfections in the corrected gradient waveforms and trajectories are attributed to asymmetrically occurring undulations in the actual, generated gradients, suggesting that the underlying assumption of linearity is violated. In conclusion, images of sufficient quality can be acquired on preclinical small-animal scanners using spiral k-space trajectories without the need to carry out separate trajectory measurements each time. Depending on the trajectory, a simple isotropic delay-correction or a GSTF-based correction can provide images of similar quality.
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction