Cardiac diffusion kurtosis imaging in the human heart in vivo using 300 mT/m gradients
Maryam Afzali1,2, Sam Coveney1, Lars Mueller1
1Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
Purpose:
Diffusion tensor imaging (DTI) is commonly used in cardiac diffusion magnetic resonance imaging (dMRI). However, the tissue's microstructure (cells, membranes, etc.) restricts the movement of the water molecules, making the spin displacements deviate from Gaussian behavior. This effect may be observed with diffusion kurtosis imaging (DKI) using sufficiently high b-values ( ), which are presently outside the realm of routine cardiac dMRI due to the limited gradient strength of clinical scanners. The Connectom scanner with enables high b-values at echo times (TE) similar to DTI on standard clinical scanners, therefore facilitating cardiac DKI in humans.
Methods:
Cardiac-gated, second-order motion-compensated dMRI was performed with in 10 healthy volunteers on a 3T MRI scanner with . The signal was fitted to a cumulant expansion up to and including the kurtosis term, and diffusion metrics such as fractional anisotropy (FA), mean diffusivity (MD), mean kurtosis (MK), axial kurtosis (AK), and radial kurtosis (RK) were calculated.
Results:
We demonstrate deviation of the signal from monoexponential decay for b-values ( ). Radial kurtosis ( ) was observed slightly larger than axial kurtosis ( ), and the difference is statistically significant ( , ).
Conclusion:
This work demonstrates the feasibility of quantifying kurtosis effect in the human heart in vivo (at an echo time shorter than typical TEs reported for cardiac DTI), using high-performance gradient systems (which are 4-8 times stronger than on standard clinical scanners). Our work lays the foundation for exploring new biomarkers in cardiac dMRI beyond DTI.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Imaging Studies for Cardiovascular System V: CT