Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
The Effects of Abnormal Exposure on Individual Dose Monitoring with TLD Dosimeters
Yanling Yi1, Michael G Stabin2
1Institute of Radiation Medicine, Fudan University, 2094 Xietu Road, Shanghai 200032, China.
Abstract:
Objectives: To analyze the effects of normal x-ray inspection, machine washing, and machine drying on thermoluminescent dosimeter (TLD) measurements during external individual monitoring and to provide suggestions for determining individual monitoring measurements under the mentioned abnormal situations. In this study, we focused on three abnormal situations: x-ray inspection, machine washing, and machine drying, which are common in external individual dose monitoring. We measured and compared the doses from TLD with and without 11, 23, 35, and 50 security checks. We used different radiation sources to expose the TLDs before or after machine washing with or without hot drying. The three radiation sources are natural background radiation, 137 Cs γ rays, and 320 kVp x-rays. We measured 20 TLDs for each situation. The average doses for the TLDs with 11, 23, 35, 50 security checks are 27.7 μGy, 59.7 μGy, 84.1 μGy, and 121.0 μGy, respectively. We measured an average dose of 2.5 μGy per exposure. The doses showed no significant difference between different times of washing with different radiation sources, natural background radiation, 137 Cs, or x-ray exposures. There was also no significant difference between the dose coming from the controlled group, drying at 60 °C and 90 °C for 1 h after exposure to 137 Cs γ rays and 320 kVp x-rays. The common machine drying under the temperature of 90 °C did not affect TLD measured doses.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
08:23An Automated Microscopic Scoring Method for the γ-H2AX Foci Assay in Human Peripheral Blood Lymphocytes
Published on: December 25, 2021