Visualization of residual gaps after linear ablation using the LUMIPOINTTM module: A case report
Yasuyuki Egami1, Yasuharu Matsunaga-Lee1, Masamichi Yano1
1Division of Cardiology, Osaka Rosai Hospital, Sakai, Osaka, Japan.
Indian Pacing and Electrophysiology Journal
|September 20, 2024
Summary
The Low Confidence Setting (L-CS) method effectively visualizes critical isthmuses in pseudo-focal atrial tachycardias (ATs). This technique aids in identifying ablation targets, leading to successful termination of ATs after linear ablation procedures.
Area of Science:
- Electrophysiology
- Cardiac Mapping
Background:
- Macroreentrant atrial tachycardias (ATs) with epicardial conduction can appear as pseudo-focal ATs on 3-D mapping.
- The Rhythmia mapping system's LUMIPOINT module highlights electrograms (EGMs) based on a confidence slider (CS).
Observation:
- Lowering the CS (L-CS) can reveal previously undetected EGMs, potentially visualizing the critical isthmus of pseudo-focal ATs.
- In three cases of pseudo-focal AT post-linear ablation, the L-CS method was applied by adjusting the LUMIPOINT activation window and gradually decreasing the CS.
- Highlighted areas before and after the ablation line expanded and fused, revealing fragmented, low-voltage EGMs at the fusion sites.
Findings:
- The L-CS method successfully visualized residual gaps and identified critical isthmuses in pseudo-focal ATs.
- Single-shot ablation at the identified sites terminated the targeted ATs in all cases.
Implications:
- The L-CS method offers a valuable tool for improving the accuracy of ablation targeting in complex atrial tachycardias.
- This technique enhances the visualization of critical isthmuses, potentially improving procedural outcomes for pseudo-focal ATs.
Related Concept Videos
Reducing Line Loss
In a three-phase circuit, line loss is an indicator of energy dissipated as heat due to the resistance of transmission lines. To address this, incorporating transformers into the system—a step-up transformer at the source and a step-down transformer at the load—is a strategic solution. Two three-phase transformers are introduced to improve this.
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
Design Example: Measuring Distance Between Two Points with Obstructions
When measuring distances in areas with physical obstructions, such as a lake in a field, surveyors must employ techniques to calculate accurate lengths without direct line measurements. One effective method is the offset technique, which allows for precise distance estimation over inaccessible stretches.In this scenario, a surveyor must measure a side of an area that crosses a lake. Since the measuring tape cannot span the lake, the surveyor begins by establishing a baseline that aligns with...