Aspiration catheter tip flutter is a reliable indicator of successful clot aspiration in ADAPT
Dan-Dong Li1,2, Jing Zheng3, Ke-Da Pan4
1Department of Neurosurgery, Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
Background:
A direct aspiration first pass technique (ADAPT) has emerged as a fast, safe, and efficacious method for treating acute large vessel occlusion. However, successful clot aspiration is not guaranteed in every ADAPT procedure. We have observed that when the catheter effectively ingested the clot, the catheter tip displayed a distinct fluttering motion, referred to herein as tip flutter. Thus this study aimed to assess whether this catheter tip flutter can be used as a sign of successful clot aspiration.
Methods:
This retrospective study included 231 consecutive patients admitted to our institution due to acute ischemic stroke and treated with ADAPT between October 2018 and November 2023. We obtained baseline and procedural data from all patients. Additionally, we assessed the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and diagnostic accuracy of the tip flutter in predicting clot aspiration.
Results:
The incidence of embolus translocation was significantly higher in the tip flutter positive group than in the tip flutter negative group (P<0.001). Also, hyperdense artery presentation was more prevalent in the positive group (P<0.001), whereas the clot burden score was higher in the negative group (P=0.002). Clot aspiration in the first pass occurred in 83 (96.5%) and 37 (25.5%) patients in the positive and negative groups, respectively (P<0.001). Multivariable logistic regression analysis showed the tip flutter sign (OR 1.09, 95% CI 0.16 to 1.29; P<0.001) was an independent predictor of successful clot aspiration. Sensitivity, specificity, PPV, NPV, and accuracy of the tip flutter for predicting clot aspiration were 69.2%, 97.3%, 96.5 %, 74.5%, and 82.7%, respectively.
Conclusions:
In this study, we found that tip flutter was a reliable indicator of successful clot aspiration during ADAPT.
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning II: Procedure