Related Experiment Video
Updated: May 7, 2026

07:54
Use of a Hanging-weight System for Isolated Renal Artery Occlusion
Published on: July 19, 2011
15.9K
Intelligent Ureteral Stent Placeable via Standard Procedure for Kidney Pressure Telemetry: An Ex-Vivo Study
Mohammad Reza Yousefi Darestani1, Dirk Lange2, Ben H Chew2
1School of Biomedical Engineering, University of British Columbia, Vancouver, Canada.
Annals of Biomedical Engineering
|September 24, 2024
Summary
Researchers developed the first wireless ureteral stent for early detection of kidney swelling (hydronephrosis) and ureteral obstruction. This "intelligent" stent uses micro-electro-mechanical systems for practical, real-time monitoring.
Area of Science:
- Biomedical Engineering
- Medical Devices
- Urology
Background:
- Ureteral stents are crucial for managing urinary tract issues.
- Early detection of hydronephrosis and ureteral obstruction is vital for patient outcomes.
- Current monitoring methods lack real-time, in-situ capabilities.
Purpose of the Study:
- To develop and test the first telemetric ureteral stent for wireless sensing.
- To enable early detection of ureteral obstruction and hydronephrosis.
- To assess the stent's compatibility with standard clinical procedures.
Main Methods:
- Prototyping a sensor-integrated ureteral stent using polymeric double-J stent and micro-electro-mechanical systems (MEMS) technology.
- Embedding micro pressure sensors and a radiofrequency antenna for wireless monitoring (40-50 MHz).
- Encapsulating the device in Parylene-C for biocompatibility and electrical insulation.
Main Results:
- The intelligent stent demonstrated wireless kidney pressure monitoring in an ex-vivo swine model.
- Ex-vivo tests showed good agreement with in-vitro results for pressure sensitivity.
- Wireless pressure tracking achieved a response of 1.3 kHz/mmHg up to 37 mmHg, covering chronic obstruction pressure ranges.
Conclusions:
- The telemetric ureteral stent is compatible with standard placement techniques.
- The device shows potential for early detection of ureteral obstruction and hydronephrosis.
- This intelligent stent offers a practical solution for real-time, wireless monitoring in clinical settings.
Related Concept Videos
Kidney Transplant II: Surgical Procedure
926
Preoperative ManagementThe primary goals of preoperative management in kidney transplantation are to optimize the patient’s metabolic state and prepare them for surgery through diet adjustments, necessary dialysis, and tailored medical treatment. This phase also involves comprehensive infection screening and patient education about the surgical procedure and postoperative care to improve outcomes and adherence.Medical ManagementA comprehensive evaluation is required for both the living...
926
Imaging Studies I: Kidney, Ureter, and Bladder Studies
871
Kidney, Ureter, and Bladder (KUB) StudiesKidney, Ureter, and Bladder (KUB) studies are standard diagnostic imaging procedures used to assess the anatomy of the urinary system. They are commonly utilized for patients experiencing abdominal pain or urinary symptoms. By using a simple X-ray of the abdomen, KUB studies can reveal structural and pathological abnormalities within the kidneys, ureters, and bladder. These studies are particularly valuable in diagnosing kidney stones, urinary...
871
Acute Kidney Injury IV: Diagnostic Studies and Prevention
559
Accurate diagnosis and effective prevention are critical in managing Acute Kidney Injury (AKI), which is linked to high mortality rates ranging from 10% to 80%. Timely recognition of at-risk patients and careful monitoring can significantly reduce the likelihood of kidney damage.Diagnostic Assessments:The diagnostic process starts with a comprehensive medical history to identify prerenal, intrarenal, and postrenal causes.Prerenal causes, such as dehydration, hypotension, or blood loss, should...
559