Accuracy assessment of artificial intelligence IOL calculation formulae: utilizing the heteroscedastic statistics and
Olga Reitblat1,2, Noa Heifetz2, Kathryn Durnford3
1Department of Ophthalmology, Rabin Medical Center, Petach Tikva, Israel.
Objectives:
To analyse the accuracy of artificial intelligence (AI)-driven intraocular (IOL) calculation formulae, together with established formulae using the heteroscedastic methodology and the Eyetemis Analysis Tool.
Methods:
Data from 404 eyes who underwent uneventful phacoemulsification with implantation of the SN60WF IOL were retrospectively reviewed. IOL power calculations were performed using the Barrett Universal II (BUII), EVO 2.0, Hoffer QST, K6, Ladas Super Formula (LSF), Nallasamy, PEARL-DGS and RBF 3.0 formulae. The SD of the prediction error (PE), served as the primary metric for accuracy. The mean absolute deviation (MAD) and the predictability rates within intervals from ±0.25 D to ±1.50 D were also evaluated. The Eyetemis Analysis Tool was used for further validation.
Results:
The SD ranged from 0.468 (Nallasamy) to 0.510 (LSF). The Nallasamy formula had a significantly lower SD than the BUII (0.505, p = 0.025) and K6 (0.489, p = 0.022) formulae. The Nallasamy formula also exhibited the lowest MAD (0.358) with a significant difference compared with the Hoffer QST formula (0.384, p < 0.001). Finally, a significantly higher percentage of eyes achieving ± 0.50 D of the target refraction was seen using the Nallasamy formula (77.19%) compared with the Hoffer QST (71.04%, p = 0.019) and Ladas Super Formula (70.79%, p = 0.030) formulae.
Conclusions:
The Nallasamy formula, incorporating AI technology, demonstrated superior accuracy according to the analysis guidelines for PE statistics for non-gaussian datasets recommended by Holladay et al. and the online Eyetemis Analysis Tool.
Related Concept Videos
Goodness-of-Fit Test
Confidence Coefficient
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Statistical Analysis: Overview
One of the most commonly used statistical quantifiers is the mean, which is the ratio between the sum of the numerical values of all results and the...
Detection of Gross Error: The Q Test
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...