Microbubble-laden aerosols improve post-nasal aerosol penetration efficiency in a preterm neonate model
Hyunhong J Min1, Xinghao Cheng1, Stephen J Payne2
1Institute of Biomedical Engineering, Department of Engineering Science, University of Oxford, Oxford, UK.
Abstract:
Nebulized lung surfactant therapy has been a neonatology long-pursued goal. Nevertheless, many clinical trials have yet to show a clear clinical efficacy of nebulized surfactant, which, in part, is due to the technical challenges of delivering aerosols to the lungs of preterm neonates. The study aimed to test microbubbles for improving lung deposition in preterm neonates. An in vitro testing method was developed to replicate the clinical environment; it used a 3D-printed preterm neonate model, connected to a high-flow nasal cannula (HFNC) and a vibrating mesh nebulizer. The flow rate of the HFNC mirrored that used in the clinics (i.e., 4, 6, and 8 L/min). Followingly, the lung penetrations of aerosols with and without microbubbles were compared. The aerodynamic diameter of aerosols with microbubbles (MMAD=1.75 μm) was lower than that of the counterpart (MMAD=2.25 μm). Microbubble-laden aerosols had a significantly higher number of microbubbles that were below 1.0 μm. Microbubble-laden aerosols had dramatically higher lung penetration in the preterm model; lung penetration efficiencies were 30.0, 25.5, and 17.5 % at 4, 6, and 8 L/min, respectively, whereas the lung penetration efficiency for conventionally nebulized aerosols was below 1.25 % in the three flow rates.