Do different types of cochlear implant electrode influence hearing preservation and speech perception?
Karin Hallin1, Ulrika Larsson1, Nadine Schart-Morén1
1Department of Surgical Sciences, Otorhinolaryngology and Head and Neck Surgery, Uppsala University, Uppsala, Sweden.
Acta Oto-Laryngologica
|October 1, 2024
Summary
Cochlear implant electrode choice impacts hearing preservation, with FLEX 24 and FLEX 20 showing superior results. Speech perception, however, remained consistent across different electrode types in this study.
Area of Science:
- Otolaryngology
- Biomedical Engineering
- Audiology
Background:
- Cochlear implantation can preserve low-frequency hearing, but optimal electrode selection for hearing preservation and speech perception remains debated.
- Understanding electrode performance is crucial for managing hearing outcomes in cochlear implant recipients.
Purpose of the Study:
- To evaluate hearing preservation and speech discrimination one year post-implantation across various cochlear implant electrode types in adult patients.
- To identify the most effective electrode for maximizing hearing preservation and maintaining speech perception.
Main Methods:
- A retrospective analysis of adult patients implanted between 2014 and 2022.
- Hearing preservation was quantified using the HEARING group formula (minimal, partial, complete).
- Speech perception was assessed via monosyllabic word recognition.
Main Results:
- FLEX 24 electrodes demonstrated statistically significant better hearing preservation compared to FLEX 28, FLEX soft, and contour advance electrodes (p < 0.05).
- FLEX 20 electrodes also showed statistically significant better hearing preservation than contour advance electrodes (p < 0.05).
- No significant differences in monosyllabic word scores were observed among the different electrode types.
Conclusions:
- Electrode type significantly influences hearing preservation in cochlear implant patients, with specific models like FLEX 24 and FLEX 20 offering advantages.
- Speech perception outcomes were not significantly affected by the electrode type used.
- Findings provide valuable data for pre-operative patient counseling regarding hearing preservation expectations with cochlear implants.