A Biomechanical Analysis of Oblique Metacarpal Metadiaphyseal Fracture Fixation in a Cadaver Model
Edgar Garcia-Lopez1, Jeremy Wafong Siu1, Jeffrey W Kwong1
1Department of Orthopedic Surgery, University of California, San Francisco, CA.
Purpose:
Although metacarpal fractures are typically managed nonoperatively, when surgical management is indicated, metacarpal fractures are commonly treated with crossed Kirschner wires (K-wires), which may limit early range of motion. Intramedullary implants are increasing in use with the potential advantage of early range of motion; however, stability in oblique metacarpal neck fractures remains a theoretical concern. The purpose of this study was to determine the biomechanical stability of noncompressive intramedullary fixation for oblique metacarpal neck fractures compared with crossed K-wire fixation.
Methods:
The index, long, and small metacarpals were harvested from three matched pairs of fresh-frozen cadavers. Oblique fractures at the metadiaphyseal region were created in each metacarpal. Each metacarpal was randomized to noncompressive, threaded intramedullary nail fixation or fixation with two crossed K-wires. Specimens were mounted in a Materials Testing System load frame and axially loaded until failure. Load to failure (LTF), stiffness, and load to 2 mm displacement were calculated from load-displacement curves. Differences in peak LTF, stiffness, and load to 2 mm displacement between noncompressive intramedullary fixation and crossed K-wire fixation were evaluated.
Results:
The noncompressive intramedullary fixation cohort had a significantly higher LTF (1,190.9 ± 534.7 N vs 297.0 ± 156.0 N) and stiffness (551.3 ± 164.6 N/mm vs 283.0 ± 194.5 N/mm) when compared with the crossed K-wire fixation cohort. Load at 2 mm displacement was greater in the noncompressive intramedullary fixation cohort compared with crossed K-wire fixation (820.5 ± 203.9 vs 514.1 ± 259.6).
Conclusions:
For oblique metadiaphyseal metacarpal fractures, noncompressive intramedullary fixation provides a biomechanically superior construct under axial loading in terms of LTF, stiffness, and load to 2 mm of displacement compared with crossed K-wire fixation.
Clinical Relevance:
Noncompressive intramedullary nails may be an alternative to K-wire fixation for the treatment of oblique metadiaphyseal metacarpal fractures.
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bones of the Upper Limb: Humerus
Bones of the Upper Limb: Ulna
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...
Bones of the Lower Limb: Femur and Patella