Dual-Column Stopped-Flow GC Method for Quantification of Permanent Gases, Water, and Organic Acids from a Single
Benjamin R Hoekstra1, Lyle E Monson1, Christopher P Nicholas1
1Låkril Technologies Corporation, Chicago, Illinois 60612, United States.
Abstract:
Analysis of permanent gases and organic acids is typically performed with separate methods due to challenges in achieving sufficient resolution when using GC columns that are chemically compatible with acids. We developed a method to analyze samples containing high levels of water, organic acids, oxygenates, and permanent gases (CO, CO2, and C2H4) using a single injection. Resolution is achieved using two columns: (1) an HP-FFAP column resistant to water and acids, which separates compounds based on hydrogen-bonding affinity, and (2) a GS-CarbonPLOT column capable of separating permanent gases. We employed a column isolation methodology using rotary valves to allow a heart cut of permanent gases, which coelute off the HP-FFAP column, to be trapped on the CarbonPLOT column. After separation and analysis of water, organic acids, and oxygenates via the HP-FFAP column, flow is resumed on the CarbonPLOT column, and the permanent gases are separated independently. By utilizing the unique combination of columns and the stopped-flow sequence, we achieved separation and quantification of the wide array of compounds occurring during lactic acid dehydration to acrylic acid from a single injection without the need for a sophisticated flow modulator or multiple GCs or detector arrays. This GC apparatus serves as an economical in-line analysis method for our catalytic studies and is applicable to a wide array of samples relevant to biobased fuels and chemicals.
More Related Videos
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...