Image quality assessments in abdominal CT: Relative importance of dose, iterative reconstruction strength and slice
B Kataria1, M Woisetschläger1, J Nilsson Althén2
1Department of Radiology, Linköping University, Linköping, Sweden; Department of Health, Medicine & Caring Sciences, Linköping University, Linköping, Sweden; Center for Medical Image Science & Visualization (CMIV), Linköping University, Linköping, Sweden.
Introduction:
Low contrast resolution in abdominal computed tomography (CT) may be negatively affected by attempts to lower patient doses. Iterative reconstruction (IR) algorithms play a key role in mitigating this problem. The reconstructed slice thickness also influences image quality. The aim was to assess the interaction and influence of patient dose, slice thickness, and IR strength on image quality in abdominal CT.
Method:
With a simultaneous acquisition, images at 42 and 98 mAs were obtained in 25 patients. Multiplanar images with slice thicknesses of 1, 2, and 3 mm and advanced modeled iterative reconstruction (ADMIRE) strengths of 3 (AD3) and 5 (AD5) were reconstructed. Four radiologists evaluated the images in a pairwise manner based on five image criteria. Ordinal logistic regression with mixed effects was used to evaluate the effect of tube load, ADMIRE strength, and slice thickness using the visual grading regression technique.
Results:
For all assessed image criteria, the regression analysis showed significantly (p < 0.001) higher image quality for AD5, but lower for tube load 42 mAs, and slice thicknesses of 1 mm and 2 mm, compared to the reference categories of AD3, 98 mAs, and 3 mm, respectively. AD5 at 2 mm was superior to AD3 at 3 mm for all image criteria studied. AD5 1 mm produced inferior image quality for liver parenchyma and overall image quality compared to AD3 3 mm. Interobserver agreement (ICC) ranged from 0.874 to 0.920.
Conclusion:
ADMIRE 5 at 2 mm slice thickness may allow for further dose reductions due to its superiority when compared to ADMIRE 3 at 3 mm slice thickness.
Implications For Practice:
Combination of thinner slices and higher ADMIRE strength facilitates imaging at low dose.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...