Icteria interference for 34 clinical chemistry analytes on different analytical platforms: Method or analyzer
Alen Vrtaric1, Marijana Miler1, Nora Nikolac Gabaj2
1Working Group for Preanalytical Phase of the Croatian Society of Medical Biochemistry and Laboratory Medicine, Zagreb, Croatia; Department of Clinical Chemistry, Sestre milosrdnice University Hospital Center, Zagreb, Croatia.
Icterus interference in clinical chemistry testing varies by analytical platform and method. Some interferences, like those affecting triglycerides, cholesterol, and urate, are more sensitive to bilirubin concentration across analyzers.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Analytical Toxicology
Background:
- Bilirubin, a common byproduct of heme metabolism, can interfere with laboratory assays.
- Icterus, characterized by elevated bilirubin levels, poses a diagnostic challenge in clinical chemistry.
- Understanding interference patterns is crucial for accurate patient results.
Purpose of the Study:
- To evaluate the impact of increasing bilirubin concentrations on 34 clinical chemistry analytes.
- To determine if icterus interference is dependent on the analytical platform and method used.
- To compare interference patterns across four different analytical platforms.
Main Methods:
- A pooled serum sample was spiked with bilirubin to create icteric samples.
- Measurements of 34 analytes were performed in duplicate on four analytical platforms.
- Bias was calculated against non-icteric samples and compared to External Quality Assurance criteria.
Main Results:
- Most analytes showed minimal interference from icterus.
- Triglycerides, cholesterol, and urate exhibited the highest sensitivity to bilirubin interference.
- Interference patterns varied significantly across different analytical platforms.
Conclusions:
- Icterus interference is not uniform across all clinical chemistry assays.
- Analyzer-specific differences in interference were observed, even with identical analytical methods.
- Method and analyzer choice are critical factors in mitigating icterus interference.