Background:
Patients born with single ventricle anatomy typically undergo surgical palliation in three stages, culminating in the Fontan procedure. Assessment of flow across a Fontan fenestration by Doppler ultrasound theoretically allows for non-invasive estimation of the transpulmonary gradient (TPG). Our objective was to determine the relationship between Doppler-derived mean fenestration gradient (mFG) and direct catheter-based measurements of TPG in patients with fenestrated Fontans.
Methods:
We performed a single-center retrospective cohort study of 59 patients with fenestrated Fontans completed between 2000 and 2022. The primary outcome was catheter-based measurement of TPG and the primary predictor was mFG from echo performed within 6 months of the catheterization. Linear regression and R2 were used to determine the relationship between predictors and outcomes.
Results:
Catheter-based measurements of TPG and mFG were weakly correlated (R2 = 0.382, p < 0.001); the regression coefficient was 0.550, with a standard error of 0.09 for every increase in mFG (Cath TPG = 0.55 [mFG] + 1.92). mFG had a slightly better predictive relationship with cath-derived TPG in patients with systemic left ventricles with R2 of 0.47, p < 0.004.
Conclusion:
mFG accounts for approximately 38% of the variance in catheter-derived TPG. Although mFG is non-invasive and intuitive, mFG in Fontan patients should be interpreted with caution and direct measurement by cardiac catheterization should be considered.
More Related Videos
04:48Noninvasive Determination of Vortex Formation Time Using Transesophageal Echocardiography During Cardiac Surgery
Published on: November 28, 2018
11:04Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
Published on: September 1, 2014
Related Concept Videos
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...