Dosimetry in MRgPT: Impact of magnetic fields on TLD dose response during proton irradiation
Hermann Fuchs1, Hugo Palmans2,3, Gerd Heilemann1
1Division of Medical Physics, Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria.
Background:
Proton beam therapy, when integrated with MRI guidance, presents complex dosimetric challenges due to interactions with magnetic fields. Prior research has emphasized the nuanced impact of magnetic fields on dosimetry. For thermoluminescent dosimeters (TLDs) the electron-return effect, alongside small air cavities surrounding the pellets, can lead to nonuniform dose distributions. Future MR-guided proton therapy will require reliable methods for end-to-end tests and dosimetric audits, which so far are often performed using TLDs equipped with phantoms. This implicates the necessity of accounting for these interactions.
Purpose:
This study investigates the influence of magnetic fields on TLDs at two proton energies, using magnetic field strengths of 0, 0.25, and , aiming to clarify their impact on dose measurement accuracy.
Methods:
The study was conducted at a synchrotron-based ion beam therapy beam line, enhanced by a resistive dipole magnet for creating magnetic fields up to to simulate MR-guided proton therapy. Individual correction factors were applied for TLD measurements. The impact of air gaps on the TLD signal was evaluated using three dedicated TLD holders with air gaps of 0.1, 0.25, and 0.5 mm surrounding the TLD pellets using the highest available proton energy of . Additionally, the influence of the magnetic field strength on the TLD response was evaluated for two proton energies of and .
Results:
The study found no statistically significant variation in TLD dose response attributable to changes in the air gap or the presence of magnetic fields. A power analysis indicated an upper limit on a potential change in dose-response as small as 1.5%.
Conclusions:
The findings suggested that the impact of air gap variations and magnetic field strengths on the TLD response was below the detection threshold of TLD sensitivity. This emphasizes the suitability of TLDs for dose measurement in MR-guided proton therapy, indicating that additional correction factors may not be necessary despite the influence of magnetic fields.
More Related Videos
07:47Custom-designed Laser-based Heating Apparatus for Triggered Release of Cisplatin from Thermosensitive Liposomes with Magnetic Resonance Image Guidance
Published on: December 13, 2015
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Magnetic Resonance Imaging
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Biological Effects of Radiation