Investigating the Effect of Blurring and Focusing Current in Cochlear Implant Users with the Panoramic ECAP Method
Charlotte Garcia1, Charlotte Morse-Fortier2, François Guérit3
1Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB27EF, UK. charlotte.garcia@mrc-cbu.cam.ac.uk.
Journal of the Association for Research in Otolaryngology : JARO
|October 16, 2024
Summary
Panoramic ECAP (PECAP) measurements show that simultaneous multi-electrode stimulation in cochlear implants (CIs) increases current spread towards the cochlear apex. Partial tripolar (pTP) stimulation yielded varied results, impacting CI optimization.
Area of Science:
- Neuroscience
- Biomedical Engineering
- Audiology
Background:
- Cochlear implants (CIs) use electrical stimulation to restore hearing.
- Optimizing CI performance involves controlling electrical current spread and neural responsiveness.
- Panoramic ECAP (PECAP) offers a method to assess these parameters along the CI electrode array.
Purpose of the Study:
- To investigate how electrical current "sharpening" (partial tripolar stimulation) and "blurring" (simultaneous multi-electrode stimulation) affect current spread and neural responsiveness.
- To evaluate the utility of PECAP in estimating these stimulation effects.
Main Methods:
- PECAP measurements were taken in 12 Advanced Bionics CI users at their most comfortable listening levels.
- Stimulation modes included standard, partial tripolar (pTP), and simultaneous multi-electrode ("blurred") stimulation.
- Current spread and neural responsiveness were estimated along the CI electrode array using PECAP data.
Main Results:
- Blurred stimulation consistently increased estimated current spread towards the cochlear apex.
- Partial tripolar stimulation showed variable effects on current spread, with a small group-level increase observed in some participants.
- PECAP successfully detected localized changes in current spread.
Conclusions:
- Simultaneous multi-electrode stimulation effectively increases current spread toward the cochlear apex, as detected by PECAP.
- Partial tripolar stimulation's effect on current spread is inconsistent.
- These findings align with speech perception data and offer insights for improving CI programming and device design.
Keywords:
Blurred stimulationCochlear implantCurrent spreadElectrically evoked compound action potential (ECAP)Focused stimulationNeural responsiveness