Comparing peripheral limb and forehead vital sign monitoring in newborn infants at birth
Suvvi K Narayana Swamy1, Simon J Stockwell1, Chong Liu1
1Optics and Photonics Research Group and Centre for Healthcare Technologies, University of Nottingham, University Park, Nottingham, UK.
Background:
To study the feasibility of measuring heart rate (HR) and oxygen saturation (SpO2) on the forehead, during newborn transition at birth, and to compare these measurements with those obtained from the wrist.
Methods:
Vital signs were measured and compared between forehead-mounted reflectance (remittance) photoplethysmography sensor (fhPPG) and a wrist-mounted pulse oximeter sensor (wrPO), from 20 enrolled term newborns born via elective caesarean section, during the first 10 min of life.
Results:
From the datasets available (n = 13), the median (IQR) sensor placement times for fhPPG, ECG and wrPO were 129 (70) s, 143 (68) s, and 159 (76) s, respectively, with data recorded for up to 10 min after birth. The success rate (percentage of total possible HR values reported once sited) of fhPPG (median = 100%) was higher compared to wrPO (median = 69%) during the first 6 min of life (P < 0.005). Both devices exhibited good HR agreement with ECG, achieving >95% agreement by 3 (fhPPG) and 4 (wrPO) min. SpO2 for fhPPG correlated with wrPO (r = 0.88), but there were significant differences in SpO2 between the two devices between 3 and 8 min (P < 0.005), with less variance observed with fhPPG SpO2.
Conclusion:
In the period of newborn transition at birth in healthy term infants, forehead measurement of vital signs was feasible and exhibited greater HR accuracy and higher estimated SpO2 values compared to wrist-sited pulse oximetry. Further investigation of forehead monitoring based on the potential benefits over peripheral monitoring is warranted.
Impact:
This study demonstrates the feasibility of continuously monitoring heart rate and oxygen saturation from an infant's forehead in the delivery room immediately after birth. Significantly higher SpO2 measurements were observed from the forehead than the wrist during the transition from foetal to newborn life. Continuous monitoring of vital signs from the forehead could become a valuable tool to improve the delivery of optimal care provided for newborns at birth.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Introduction to Vital Signs
Vital signs help healthcare professionals assess an individual's well-being and detect any functional changes...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.