Related Experiment Video
Updated: Jun 10, 2025

How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Reference values for cerebral and peripheral tissue oximetry in children: A clinical TD-NIRS study
Valeria Calcaterra1,2, Michele Lacerenza3, Virginia Rossi2
1Pediatric and Adolescent Unit, Department of Internal Medicine, University of Pavia, Pavia, Italy.
Aim:
Current non-invasive near-infrared spectroscopy (NIRS) tissue oximetry suffers from suboptimal reproducibility over probe repositioning, hindering clinical threshold establishment. Time Domain-NIRS (TD-NIRS) offers higher precision but lacks sufficient paediatric data, preventing effective clinical application. We aimed to establish reference ranges for cerebral and mid-upper arm (MUA) tissue haemodynamics in paediatric subjects using TD-NIRS and explore correlations with auxological variables.
Methods:
TD-NIRS measurements were conducted acquiring data from cerebral and MUA regions with the NIRSBOX tissue oximeter. Morphological and clinically relevant information were collected to explore potential correlations with TD-NIRS derived parameters.
Results:
TD-NIRS assessment was applied in 350 children (8.4 ± 5.0 years). Precision of TD-NIRS was demonstrated with standard deviations of 0.9% (StO2) and 4.2 μM (tHb) for frontotemporal cerebral cortex, and 0.8% (StO2) and 3.7 μM (tHb) for MUA. No user dependency was observed. The trends of values for cerebral and peripheral regions vary differently according to age and auxological parameters.
Conclusion:
This study reports resting-state optical and haemodynamic values for a healthy paediatric population, providing a foundation for future investigations into clinically relevant deviations in these parameters. Furthermore, correlations with anthropometric and demographic values provide valuable insights for a deeper understanding of tissue haemodynamic evolution in childhood.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation