Performance evaluation of a small-animal PET scanner with 213 mm axis using NEMA NU 4-2008
Qing Wei1,2,3,4, Yingjie Wang1,3,4, Xianchao Huang1,3,4
1Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing, China.
Background:
Long-axis positron emission tomography (PET) has emerged as one of the recent research directions in PET due to its ability to significantly enhance sensitivity and counting performance for low-dose imaging, rapid imaging, and whole-body dynamic imaging.
Purpose:
The PET system presented in this study is a long-axis animal PET based on lutetium-yttrium orthosilicate and silicon photomultiplier, designed for whole-body imaging in rats. It features a diameter of 143 mm and an axial length of 213.3 mm. This study evaluated the performance of this PET system in accordance with the National Electrical Manufacturers Association (NEMA) NU 4-2008 standards.
Methods:
The performance evaluation was conducted according to the NEMA NU 4-2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction (SF) and image quality. In addition, a rat imaging study was conducted to assess the imaging capability of this PET system.
Results:
The average energy resolution of the PET system was 12.87%, the average coincidence timing resolution was 751 ps. The FWHM of spatial resolution reconstructed by filtered back projection and 3D-OSEM-PSF algorithm at 5 mm radial offset from the axial center were 1.65 and 0.88 mm. The peak absolute sensitivity measured by a point source at the center of the field of view was evaluated as 6.71% (361-661 keV) and 10.31% (250-750 keV). For the mouse-like phantom, the SF was 11.0% and the peak noise equivalent counting rate (NECR) was 1193 kcps at 94.2 MBq (2.54 mCi). For the rat-like phantom, the SF was 26.8% and the NECR was 682.5 kcps at 78.6 MBq (2.12 mCi).
Conclusions:
The performance measurement results demonstrate that this PET system exhibits high sensitivity and count rate performance, making it potential for high-quality whole-body dynamic imaging of rats.