Related Experiment Video
Updated: Jul 30, 2026

Rapid Homogeneous Detection of Biological Assays Using Magnetic Modulation Biosensing System
Published on: June 13, 2010
Ratiometric Electrochemical Detection of Interleukin-6 Using Electropolymerized Methylene Blue and a Multi-Walled
Zhuo Liu1, Fengyu Liu2, Chaofan Wang1
1College of Chemistry & Pharmacy, Northwest A&F University, Xianyang 712100, China.
Abstract:
Herein, we report a ratio-based electrochemical biosensor for the detection of interleukin-6 (IL-6). We electropolymerized methylene blue (MB) on the surface of screen-printed carbon electrodes; introduced an internal reference signal probe; modified the carboxylate multi-walled carbon nanotubes on the electrode surface to increase the electrochemically active area; and finally linked the amino-modified IL-6 aptamer to the electrode surface through the Schiff base reaction, with bovine serum albumin (BSA) added to mask non-specific adsorption. After adding IL-6 to the samples, the signal of IMB remained almost unchanged, while the signal of I[Fe(CN)6]3-/4- decreased with increasing IL-6 concentration. Thus, a novel ratiometric electrochemical sensor with a linear range of 0.001~1000.0 ng/mL and a low detection limit of 0.54 pg/mL was successfully developed. The sensor had high repeatability, stability, sensitivity, and practicability. It provides a new method for constructing proportional electrochemical sensors and detecting IL-6.