Design and Use of a Custom Phantom for Regular Tests of Radiography Apparatus: A Feasibility Study
Nikolay Dukov1, Vanessa-Mery Valkova2, Mariana Yordanova3
1Department of Medical Equipment, Electronic and Information Technologies in Healthcare, Medical University of Varna, 9002 Varna, Bulgaria.
Abstract:
This study investigates the feasibility of employing an in-house-developed physical phantom dedicated to the weekly quality control testing of radiographic systems, performed by radiographers. For this purpose, a 3D phantom was fabricated, featuring test objects, including a model representing a lesion. Alongside this phantom, a commercial phantom, specifically, IBA's Primus L, was utilized. Weekly imaging of both phantoms was conducted over a span of four weeks, involving different imaging protocols and anode voltages. Subsequently, the obtained data underwent visual evaluation, as well as measurement of the intensity of selected regions of interest. The average values for three incident kilovoltages remained consistently stable over the four weeks, with the exception of the "low energy" case, which exhibited variability during the first week of measurements. Following experiments in "Week 1", the X-Ray unit was identified as malfunctioning and underwent necessary repairs. The in-house-developed phantom demonstrated its utility in assessing the performance of the X-Ray system.