MRI-quality and morphometric MRI analysis to identify focal cortical dysplasia: An exploratory study
E N Zuidhoek1, J N P Zwemmer1, G H Visser1
1Stichting Epilepsie Instellingen Nederland (SEIN), Department of Clinical Neurophysiology, Heemstede, the Netherlands.
Reduced MRI quality, including signal-to-noise ratio (SNR) and spatial resolution (SR), significantly impacts the detection of focal cortical dysplasia (FCD) using Morphometric Analysis Program (MAP18). Lower quality scans can render FCD undetectable, potentially affecting surgical outcomes for epilepsy patients.
Area of Science:
- Neuroimaging
- Epilepsy Research
- Medical Image Analysis
Background:
- Morphometric Analysis Program v2018 (MAP18) assists in detecting subtle focal cortical dysplasia (FCD) in normal MRI scans for epilepsy surgery evaluation.
- Investigated the impact of reduced MRI signal-to-noise ratio (SNR) and spatial resolution (SR) on MAP18's FCD detection capabilities.
Purpose of the Study:
- To assess how diminished MRI quality (SNR and SR) affects MAP18's ability to detect focal cortical dysplasia (FCD).
- To identify potential limitations of MAP18 in identifying FCD under suboptimal imaging conditions.
- To inform strategies for improving surgical planning and patient outcomes in focal epilepsy.
Main Methods:
- Retrospective analysis of 30 MRI scans with radiologically confirmed FCD.
- Artificial reduction of SNR and SR in ten incremental steps.
- Multilevel analysis to evaluate the impact of reduced SNR and SR on MAP18 detection metrics (z-score, volume, cluster count).
Main Results:
- Reduced SR and SNR significantly affected FCD cluster z-score, volume, and the number and volume of detected clusters.
- FCD became undetectable by MAP18 after six SR reduction steps (2.8 × 2.8 × 2.8 mm³ voxel size) and two SNR reduction steps.
- SNR reduction also significantly impacted the z-score of the total detected cluster count.
Conclusions:
- Reduced spatial resolution and signal-to-noise ratio negatively impact FCD detection using MAP18 post-processing.
- MAP18 evaluators must assess MRI quality, noting that voxel sizes of 2.8 × 2.8 × 2.8 mm³ and higher, or significant visual noise, may impede FCD detection.
- Ensuring high-quality MRI scans is crucial for accurate FCD detection and successful epilepsy surgery outcomes, as repeating low-quality scans is preferable to missed diagnoses.
More Related Videos
09:06Whole-brain Segmentation and Change-point Analysis of Anatomical Brain MRI—Application in Premanifest Huntington's Disease
Published on: June 9, 2018
11:50A Standardized Pipeline for Examining Human Cerebellar Grey Matter Morphometry using Structural Magnetic Resonance Imaging
Published on: February 4, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging