Different Positions of Weightbearing CT Images Can Influence the Hindfoot Alignment Evaluation Using 2-Dimensional
Mingjie Zhu1, Wanjun Gu2, Chao Wang3
1Department of Orthopedic Surgery, University of Colorado School of Medicine, Denver, CO, USA.
Background:
Three-dimensional weightbearing CT (WBCT) has been widely used to assess foot and ankle alignment. However, most current measurement methods are based on 2-dimensional concepts-distance, long axis, angulation, etc-and are sensitive to changes in orientation of the WBCT images. This study demonstrated how changes in positioning of the image can significantly influence the evaluation of hindfoot alignment.
Methods:
WBCT scans of 10 feet without deformities were used. In the horizontal view, the long axis of the foot image was aligned to a neutral position and then rotated 5, 10, 15, and 20 degrees both internally and externally to simulate malposition. The Calcaneal Moment Arm (CMA) was measured by 2 investigators independently in the above positions. An intraclass correlation coefficient (ICC) model was used to assess the intra- and interobserver reliabilities. The correlation between the CMA and the rotation angle of the foot images was analyzed using linear regression.
Results:
The CMA measurements demonstrated that internal rotation of the image changed the neutral hindfoot into valgus, whereas external rotation led to varus. A 1-degree internal or external rotation of the image correlated with 0.48 (±0.03) mm increase or decrease in the modified WBCT-CMA value (P < .0001, R2 = 0.6406).
Conclusion:
This study demonstrated that although 3D WBCT provides the ability to observe the foot from different perspectives, current alignment evaluation tools are limited to 2 dimensions. Therefore, positioning the WBCT images in a consistent orientation is important to generate correct data.