A multicenter study comparing the bacterial reduction on flexible endoscopes without a working channel between UV-C
Yana Halmans1,2, D J Wellenstein3, M Romijn3
1Department of Otorhinolaryngology and Head and Neck Surgery, Radboud University Medical Center Nijmegen, P/O Box 9101, Nijmegen, The Netherlands. yana.halmans@radboudumc.nl.
Background:
To prevent cross-contamination between patients, adequate reprocessing is necessary when using flexible endoscopes (FEs) without a working channel. The current reprocessing process using an Endoscope Washer Disinfector (EWD) is time-consuming. Ultraviolet light group C (UV-C) exposition is an alternative and fast disinfection method and has previously been shown to adequately reduce Colony Forming Units (CFUs) on FEs without a working channel. The objective of this study was to examine whether UV-C light is as effective in reducing CFUs on contaminated FEs without a working channel compared to the EWD.
Methods:
FEs without a working channel were collected in three different Otorhinolaryngology Departments in the Netherlands. After pharyngolaryngoscopy, a manual pre-cleaning with tap water was performed and a culture was collected by rolling the distal 8-10 cm of the FE over an agar plate. Next, the FE was randomly assigned to be disinfected with UV-C light (D60) or the EWD (gold standard). After disinfection, another culture was taken. The primary outcome was microbiological contamination, defined by Colony Forming Units (CFU).
Results:
600 FEs without a working channel were randomized. After clinical use and manual pre-cleaning, 239/300 (79.7%) FEs in the UV-C group and 262/300 (87.3%) FEs in the EWD group were contaminated (i.e., > 0 CFU). FEs without culture confirmed contamination were excluded from further analysis. After UV-C light disinfection, 195/239 (81.6%) FEs showed 0 CFUs, compared to 187/262 (71.4%) FEs disinfected with the EWD (p < 0.01). A multivariate logistics regression analysis showed an increased odds of 0 CFUs when using UV-C light (OR 1.83, 95% CI 1.19-2.79; p < 0.01), conditional on participating hospitals and types of FE.
Conclusions:
UV-C light disinfection of FEs without a working channel appears more effective in reducing CFUs compared to the EWD and might be a good alternative disinfection method.
Trial Registration:
Not applicable.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods for Controlling Microbial Growth