Related Experiment Video
Updated: Jun 17, 2026

07:21
Clinical Application of 24 G Cannula Needle and 3-0 Polypropylene Suture in Vas Deferens Exploration
Published on: February 10, 2023
2.6K
Optimal deflection techniques for flexible and navigable suction ureteral access sheaths (FANS): a comparative in
Alex Lua1, Lynnette R L Tan1, Frédéric Panthier2,3
1Department of Urology, Tan Tock Seng Hospital, 11 Jalan Tan Tock Seng, Singapore, 308433, Singapore.
World Journal of Urology
|October 30, 2024
Summary
Optimizing flexible and navigable ureteral access sheath (FANS) deflection is crucial for urologic procedures. The sheath advancement technique and specific sheath-ureteroscope combinations yield superior angles, aiding lower pole access.
Area of Science:
- Urology
- Minimally Invasive Surgery
- Medical Device Engineering
Background:
- Flexible and navigable ureteral access sheaths (FANS) are essential tools in modern urology.
- Current evidence lacks clear guidelines on optimizing deflection capabilities of FANS.
- Understanding deflection mechanics is key to improving procedural efficiency and outcomes.
Purpose of the Study:
- To evaluate in vitro deflection angles achieved with two distinct FANS manipulation techniques: sheath advancement and ureteroscope deflection.
- To assess the impact of various sheath size-ureteroscope combinations on FANS deflection.
- To provide evidence-based insights for optimizing FANS usage in clinical practice.
Main Methods:
- In vitro evaluation of deflection angles for 10/12Fr, 11/13Fr, and 12/14Fr FANS.
- Testing with six different single-use flexible ureteroscopes (7.5Fr to 9.5Fr).
- Comparison of two deflection techniques: sheath advancement over a deflected ureteroscope and ureteroscope deflection relative to the sheath tip.
Main Results:
- Significant differences in maximum deflection angles were observed across ureteroscopes and techniques (ANOVA p < 0.01).
- The sheath advancement technique yielded the largest angles (218°-277°), exceeding manufacturer specifications in most cases.
- 10/12Fr and 11/13Fr sheaths demonstrated greater flexibility; the 11/13Fr sheath with OTU8.6Fr/Innovex8.7Fr ureteroscopes provided the largest angles.
Conclusions:
- Optimal FANS deflection is achievable via sheath advancement or ureteroscope deflection at the tip.
- Manufacturer-specified deflection angles appear difficult to attain even with optimized techniques.
- The sheath advancement technique and 11/13Fr sheath-ureteroscope combinations are recommended for challenging lower pole access, guiding urologists' FANS technique.