Potential dose reduction and image quality improvement in chest CT with a photon-counting CT compared to a new
Joël Greffier1, Djamel Dabli1, Sebastian Faby2
1IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, Nîmes, France.
Purpose:
To compare potential dose reduction and quality improvement in chest CT images with Photon-Counting CT (PCCT) versus a Dual-Source CT (DSCT).
Methods:
Acquisitions on phantoms were performed on a DSCT and a PCCT at 5 dose levels (9.5/7.5/6.0/2.5/0.4 mGy). Noise power spectrum (NPS) and task-based transfer function (TTF) were calculated to assess noise magnitude and noise texture (fav) and spatial resolution (f50), respectively. Computed detectability indexes (d') modelled the detection of 2 chest lesions: subsolid pulmonary nodules (SPN) and high-contrast pulmonary nodules (HCN). Two radiologists subjectively assessed the quality of chest images on an anthropomorphic phantom.
Results:
For all dose levels, noise magnitude was significantly lower with PCCT than with DSCT (-44.7 ± 3.0 %; p < 0.05). Identical outcomes were found for noise texture (fav; -6.2 ± 0.5 %; p < 0.05). f50 values were significantly higher with DSCT than with PCCT from 9.5 to 6 mGy for iodine insert (p < 0.05) and from 7.5 to 2.5 mGy for air insert (p < 0.05), but similar for both inserts at other dose levels. For all dose levels, d' values were significantly higher with PCCT than DSCT (71.9 ± 5.4 % for HCN and 65.6 ± 13.5 % for SPN). From 9.5 to 2.5 mGy, the potential dose reduction was -59.0 ± 3.9 % for both lesions with PCCT compared to DSCT. Chest images were rated satisfactory for clinical use by the radiologists with both CTs for all dose levels, except at 0.4 mGy.
Conclusion:
Noise magnitude and detectability of chest lesions were better with PCCT than with the DSCT. PCCT may offer great potential for dose reduction in patients undergoing chest CT examinations.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
X-ray Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...