Diagnostic accuracy of the point-of-care standard G6PD test™ (SD Biosensor) for glucose-6-phosphate dehydrogenase
Juan Camilo Martínez1, Viviana Vélez-Marín1, Mary Lopez-Perez2
1Unit of Evidence and Deliberation for Decision Making UNED, Medical Research Institute, School of Medicine, University of Antioquia, Medellin, Colombia.
Background:
Glucose-6-Phosphate Dehydrogenase deficiency (G6PDd) is a common genetic enzymopathy that can induce haemolysis triggered by various factors, including some anti-malarial drugs. Although many Point-of-Care (PoC) tests, such as Standard G6PD™ are available to detect G6PDd, its pooled diagnostic test accuracy (DTA) remains unknown.
Methods:
To estimate the DTA of StandG6PD-BS at various thresholds of G6PDd, a systematic review with a DTA meta-analysis were conducted, searching EMBASE, MEDLINE, and SciELO databases up to April 4, 2024.The included studies were those that measured G6PD activity using StandG6PD-BS (reference test) and spectrophotometry (gold standard) in patients suspected of having G6PDd. The risk of bias (RoB) of the studies was assessed using the QUADAS-2 tool and the certainty of evidence (CoE) with the GRADE approach. For the estimation of within-study DTA, a random-effect bivariate meta-analysis was performed to determine the pooled sensitivity and specificity for 30%, 70%, and 80% enzyme levels' thresholds, and a graphical analysis of the heterogeneity using crosshair and Confidence Regions on receiver operating characteristic (ROC) space plots.
Results:
After screening 2496 reports, four studies were included with 7864 participants covering all thresholds. Two studies had high RoB in QUADAS-2 domains 2 and 3, and the others had low RoB, with low, moderate, and high heterogeneity at the 30%, 70%, and 80% thresholds, respectively. The pooled sensitivity was 99.1%, 95.7%, and 90% for 30%, 70%, and 80% thresholds, respectively. The pooled specificity was 97.4%; 92.9%; and 89.0% for 30%, 70%, and 80% thresholds, respectively.
Conclusion:
StandG6PD-BS is a PoC test with high sensitivity and specificity to detect G6PDd at different thresholds.